MET ONE 3411 Portable with Gas Calibration

2088800-05


Part of the MET ONE Facility Monitoring System. MET ONE 3411 Portable Air Particle Counter
 

Caractéristiques

Platform APC Instruments
Performance ISO 21501-4 unit-to-unit accuracy & reproducibility
Industry Standards ISO 21501-4 , 21 CFR Part 11, ISO 14644-1:2015, FCC, CE
Light Source Helium neon laser
Particle Size Analysis Range 0.1 - 5 μm / 0.1 - 5 μm
Coincidence Loss % at 1,765,000 particles/m3 (per ISO 21501-4 method)
Sample Volume 5 µm
Throughput 28.32 L per minute
Interfaces & Protocols RS232, RS-485, USB Client (v1.1), USB Host (v1.1), Modbus TCP, Modbus RTU, Serial FX, Ethernet 802.3
Data Storage 5000 samples
Reporting EU GMP (2015), ISO 146441-1.2015, FS209E, BS5295, ISO 14644-1:1999
Battery Lithium ion smart battery (rechargeable, ejectable & hot-swappable)
Battery Operating Time 3 hours (minimum)
Battery Recharge Time 6.75 hours (minimum) - 10 hours (maximum)
Pump Type Air vacuum, rated for continuous use
Zero Count ISO 21501-4 & JIS B 9921 (1 count or less in 5 minutes; 95% Upper Confidence Limit)
Counting Efficiency 50% at 0.1 μm (100% for particles > 0.15 μm per ISO 21501-4 & JIS B 9921)
Count Alarms, Display & Cycles User-defined alarm counts, Raw counts: N/CF, N/CM, N/L in cumulative or differential mode
Label Locations ID: 0 to 999; NAME: Alphanumeric (appears on printout)
Environmental Conditions Indoor Environment
Temperature Range Operating temperature range: 0°C to 40°C, Storage temperature range: -40°C to 50°C
Humidity Restrictions 10%-90% relative humidity (non-condensing)
Power Requirements 24 VDC; 3.2A with 100-240 VAC; 50/60 Hz adapter (included in ship kit)
Operating System Windows® CE
Software Simply Paperless Secure PDF Option
Resolution 320 x 240
Display Color 1/4 VGA TFT touch screen
Operating Modes Semi-automated, manual, controlled by operator
Autoclavable No
Nonpyrogenic Yes
Weight 15.9 kg (without battery)
Width 33 cm / 13 in
Depth 56 cm / 22 in
Height 23 cm / 9.1 in
Color Silver
Package Quantity 1

Customers Also Viewed

TOUS LES PRODUITS NE SONT PAS DISPONIBLES DANS TOUS LES PAYS
LA DISPONIBILITE DU PRODUIT ET LE STATUT REGLEMENTAIRE DEPENDENT DE L’ENREGISTREMENT SELON LE REGLEMENT APPLICABLE PAR PAYS