Avanti® J-26S XP Non-IVD Centrifuge, 50/60 Hz, 200/208/240V

B14535


Redesigned to provide even more versatility and safety. Built upon standard-setting performance, including high speed, low heat output, imbalance tolerance and low noise, the J-26S XP’s vast range of separations only continues to grow. Above and beyond its BioSafety systems, this centrifuge has a more impact-resistant lid, full containment barrier and automatic rotor identification. Its ergonomic design--with lower instrument height and hands-free door operation--remains unchanged. The J-265 XP can process six liters of fermentor output in less than 10 minutes, deliver a subcellular fraction at 82,000 x g and run 24 microplates at a time. It automatically detects which rotor is installed and prevents the rotor from running above its maximum-rated speed, and enables quicker, quieter, more energy-efficient runs with better temperature control. The exclusive high-torque, switched reluctance (SR) drive shortens cycle times.

Caractéristiques

Platform Avanti
Maximum Speed 26,000 rpm
BioSafe No
Acceleration Maximum or Slow
Deceleration Maximum, slow, or off
Speed Control ± 10 rpm of set speed or 0.1%, whichever is greater
Timed Run 1 minute-3 hours; Continuous Hold
Max Capacity Volume Range 1 - 6 L
Refrigeration System Non-CFC, non-ozone depleting refrigerant
Temperature Range(s) 10°C - 40°C in 1 degree increments
Rotor ID Automatic
Zonal Capability Yes
Pollution Degree 2
Environmental Conditions -10°C to 40°C in 1° increments
Software Virtual Network Computing (VNC), Mobile Fuge app for Apple iOS and Android
Material(s) Steel, Plastic
Finish Powder Coat Paint
Weight 290 kg / 639.3 lb
Width 71 cm / 28 in
Depth 86.0, 102.0 with air direviter extending cm / 86.0, 102.0 with air direviter extending in
Height 86 cm (149 cm with door open)
Clearances 7.6 cm (sides) & 16 cm (rear)
Humidity Restrictions < 95% (non-condensing)
Power Requirements 200/208/240 V, 50/60 Hz

TOUS LES PRODUITS NE SONT PAS DISPONIBLES DANS TOUS LES PAYS